The size of the touchscreen display alone speaks for itself: from the operating field, the surgeon always has a clear view of all control elements. This results in less setting adjustment or modifications. Recommended connection Based on pre-programmed settings the stepGUIDE supports you in selecting a socket for your chosen instrument. Slot assignment The active slot and the instrument in use are shown on the display and through the illuminated socket frame.

Author:Kazralkree Kigashakar
Country:Chad
Language:English (Spanish)
Genre:Music
Published (Last):25 May 2019
Pages:163
PDF File Size:1.32 Mb
ePub File Size:8.48 Mb
ISBN:716-5-82047-140-2
Downloads:55259
Price:Free* [*Free Regsitration Required]
Uploader:Shaktigami



APC operating principle Argon plasma coagulation, APC for short, is an electrosurgical, monopolar procedure for superficial hemostasis, devitalization and ablation using ionized argon gas. Argon is an inert gas and can be easily ionized. The distance between the applicator tip and tissue during use is approx. The coagulation is directly visible on the tissue; movement can be used to determine and easily control the surface distribution and the intensity of the coagulation surface.

In literature, usage is sometimes compared to using a paintbrush. Only use where there is good visibility APC should only be applied where there is good visibility Penetration depth and dosage The penetration depth of APC depends on various factors such as mode, effect and duration of application.

In the case of conventional usage, a penetration depth of approx. Avoid tissue contact Do not place the APCapplicator tip on tissue while activated. Do not allow the gas to enter the vessels or organs. Distance to metal and surgical suture To prevent unintended coagulation, do not apply APC to metal clips or surgical suture.

With or without argon support as desired.